Wika DG-10 Manuel d'utilisateur Page 14

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 112
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 13
14 WIKA operating instructions digital pressure gauge, model DG-10
11250313.03 09/2012 GB/D/F/E
GB
6. Commissioning, operation
6. Commissioning, operation
CAUTION!
Prior to commissioning, the digital pressure gauge must be subjected to a visual inspection.
Only use the digital pressure gauge if it is in perfect condition with respect to safety.
Required tool: Open-ended spanner (spanner width 27)
6.1 Mechanical mounting
The sealing faces at the instrument have to be
undamaged and clean.
When screwing the instrument in, the force
required to do this must not be applied through
the case, but only through the spanner ats
provided for this purpose and using a suitable
tool.
The correct torque depends on the dimensions
of the process connection and the gasket used
(form/material).
Do not overwind the process connection, do not exceed the permissible torques.
When screwing in, do not cross the threads.
For information on tapped holes and welding sockets, see Technical information IN 00.14
at www.wika.com.
Spanner ats
Vue de la page 13
1 2 ... 9 10 11 12 13 14 15 16 17 18 19 ... 111 112

Commentaires sur ces manuels

Pas de commentaire