Valve 3, 6 and 7 are closed
Valves 2, 4, A and B are open.
Then carry out adjustment, see chapter "Set parameters".
8.4 Density and interface measurement
For density and interface measurements, DPT10 with bilateral isolating
diaphragm CSB is used.
DPT10 in this version is immediately ready for operation.
8.5 Dierential pressure measurement
For dierential pressure measurements, DPT10 without isolating
diaphragm or with bilateral isolating diaphragm CSB is used.
DPT10 with bilateral isolating diaphragm CSB is immediately ready for
operation.
Before adjusting DPT10 without isolating diaphragm, the eective
pressure lines must be cleaned and the instrument lled with medium.
+
I
II
+
6
7
3
24
AB
Fig. 49: Preferred installation for gases
I DPT10
II 3-fold valve block
2, 4 Inlet valves
3 Breather valve
6, 7 Vent valves on DPT10
A, B Blocking valves
Instructions
Installation for gases
66 Dierential pressure transmitter DPT10 - metal measuring diaphragm Probus PA
8 Set up
37245-EN-100120