14028755.02 08/2012 GB/D/F/E
WIKA operating instructions model TGT73 19
8. Electrical connection / 9. Maintenance ...
2-wire system 3-wire system
e.g. 4 ... 20 mA e.g. 0 ... 10 V
Earth, connected
to case
Terminals 3, 4, 5 and 6:
only for internal use
U
B
+/Sig+
+0V/Sig-
Evaluation
(display)
Power supply
Terminal box
Earth, connected
to case
I
out
/U
out
Terminals 4, 5 and 6:
only for internal use
U
B
+/Sig+
+0V/Sig-
Evaluation
(display)
Power supply
As a power supply, an unstabilised DC power supply is sucient, with
a max. ripple of 10 % ss in the range of the specied power supply
limits. It is important to ensure that the applied power supply is at least
greater than the maximum required voltage drop from the external
display and evaluation units; i.e. the voltage at the transmitter must
not fall below 12 V.
9. Maintenance and cleaning
9.1 Maintenance
Gas-actuated thermometers with transmitters are maintenance-free!
The indicator should be checked once or twice every year. To do this
the instrument must be disconnected from the process and checked
using a temperature calibrator.
Repairs must only be carried out by the manufacturer.