Wika DPT-10 Manuel d'utilisateur Page 64

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 92
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 63
+
A
B
6
7
3
2
4
1
5
+
II
I
IIIIII
IV
Fig. 48: Preferred installation for closed vessels with steam overlay
I DPT10
II 3-fold valve block
III Precipitator
IV Condensate vessel
1, 5 Drain valves
2, 4 Inlet valves
3 Breather valve
6, 7 Vent valves on DPT10
A, B Blocking valves
Proceed as follows:
1 Fill the vessel to just above the lower tap
2 Fill measuring system with medium
Open valve A and B: Open block valves
Fill the minus eective pressure line on the height of the
condensation pot
3 Remove air from instrument:
Open valve 2 and 4: Discharge medium
Open valve 3: Equalisation plus and minus side
Briey open valve 6 and 7, then close again: Fill the measuring
instrument completely with the medium and remove air
4 Put measurement loop into operation:
Close valve 3: Separate plus and minus side
Open valve 4: Connect minus side
Installation for closed
vessels with steam layer
Prepare the adjustment
64 Dierential pressure transmitter DPT10 - metal measuring diaphragm F oundation Fieldbus
8 Set up
37246-EN-100120
Vue de la page 63
1 2 ... 59 60 61 62 63 64 65 66 67 68 69 ... 91 92

Commentaires sur ces manuels

Pas de commentaire