7
1.13 Resource Block Parameter List, Fieldbus Foundation
Parameter
Rel.
Index
Description Type Store
Size
byte
RO /
R/W
Min. Max. Default
ST_REV
1
Is incremented each time that there is a change in a static
parameter in the physical block.
Un-
signed
16
SRC 2 RO 0
TAG_DESC
2
Tag name of the block. This parameter must be unique in the
configuration.
OCTET_
STRING
SRC 32 R/W »«
STRATEGY
3
This can be used to group a Function Block. It is a user sup
-
plied parameter for identification purpose.
Un-
signed
16
SRC 2 R/W 0
ALERT_KEY
4 Alert keys
Un-
signed
8
SRC 1 R/W 0
MODE_BLK
5 Block running mode DS-69 Mix 4 *
1,1,
17,16
BLOCK_ERR
6 Block errors
BIT_
STRING
D 2 RO 0
RS_STATE
7 State of the function block application state machine
Un-
signed
8
D 1 RO 0
TEST_RW
8 Read/write test parameter used only for conformance testing DS-85 D 112 R/W 0..0
DD_RESOURCE
9
String identifying the tag of the resource which contains the
Device Description for this resource.
VISIBLE_
STRING
SRC 32 RO » »
MANUFAC_ID
10
Enumeration; controlled by FF
Manufacturer identification number - used by an interface
device to locate the DD file for the resource.
Un-
signed
32
SRC 4 RO
WIKA
DEV_TYPE
11
Manufacturer’s model number associated with the resource -
used by interface devices to locate the DD file for the resource.
Un-
signed
16
SRC 2 RO
0x00ec
DEV_REV
12
Manufacturer revision number associated with the resource
- used by an interface device to locate the DD file for the
resource.
Un-
signed
8
SRC 1 RO
0x03
DD_REV
13
Revision of the DD associated with the resource - used by an
interface device to locate the DD file for the resource.
Un-
signed
8
SRC 1 RO
0x01
GRANT_DENY
14
Access Permissions. Options for controlling access of host
computer and local control panels to operating, tuning and
alarm parameters of the block.
DS-70 SRC 2 R/W 0
HARD_TYPES
15 The types of hardware available as channel numbers.
BIT_
STRING
SRC 2 RO 0
RESTART
16
1: Run,
2: Restart resource,
3: Restart with defaults,
4: Restart processor
Allows a manual restart to be initiated. Several degrees of
restart are possible.
Un-
signed
8
D 1 R/W 1
FEATURES
17 Used to show supported resource block options.
BIT_
STRING
SRC 2 RO 0
FEATURE_SEL
18 Used to select resource block options.
BIT_
STRING
SRC 2 RW 0
CYCLE_TYPE
19
Identifies the block execution methods available for this
resource
BIT_
STRING
SRC 2 RO 0xC000
CYCLE_SEL
20 Used to select the block execution method for this resource.
BIT_
STRING
SRC 2 ** 0xC000
MIN_CYLCE_T
21
Time duration of the shortest cycle interval of which the
resource is capable.
Un-
signed
32
SRC 4 RO 0
MEMORY_SIZE
22
Available configuration memory in the empty resource. To be
checked before attempting a download.
Un-
signed
16
SRC 2 RO 0
NV_CYCLE_T
23
Interval between writing copies of NV parameters to non-vola
-
tile memory. Zero means never.
Un-
signed
32
SRC 4 RO 0
FREE_SPACE
24
Percent of memory available for further configuration. Zero in a
preconfigured resource.
Floating
Point
D 4 RO 0.0
FREE_TIME
25
Percent of the block processing time that is free to process
additional blocks.
Floating
Point
D 4 RO 0.0
SHED_RCAS
26
Time duration at which to give up on computer writes to func
-
tion block RCas locations.
Un-
signed
32
SRC 4 R/W 640000
SHED_ROUT
27
ms time duration at which to give up on computer writes to
function block ROut locations.
Un-
signed
32
SRC 4 R/W 640000
Vue de la page 6