20 WIKA operating instructions ProcessCalibrator, model CPH6000
11069023.04 11/2012 GB/D
GB
Voltage
supply
+
-
Example with a WIKA pressure transmitter with V signal as the test item:
1. without its own power supply, DC 24 V must be activated via the menu (see chapter
7.3 "Operating modes")
2. with its own power supply available
4.4 CPT6000 reference pressure sensor
For the CPH6000 ProcessCalibrator, there are many reference pressure sensors
to choose from (with measuring ranges from 250 mbar up to 1,000 bar with an
accuracy of 0.025 %; and measuring ranges over 1,000 bar up to 6,000 bar with an
accuracy of 0.1 %) which can be interchanged quickly and without tools. When the
ProcessCalibrator is switched on, the reference pressure sensor attached is recognised
automatically, so that no further conguration is needed.
4. Design and function