Automated Pressure Calibrator
CPC 6000
56 Mensor/WIKA Operating Instructions - CPC 6000
94 Rsetpt? <sp>{value}<cr><lf> Returns the current rate setpoint of the active
channel in current units per second.
95
Sbaud {9600, 19200, 38400, 57600} Sets the serial baud rate.
96 Sbaud? <sp>{value}<cr><lf> Returns the serial baud rate.
97
Sdata {7 or 8} Sets the serial data bits.
98 Sdata? <sp>{value}<cr><lf> Returns the serial data bits number.
99 Sensor C, X Sets the active sensor where C = Primary or
Secondary and X is the turndown.
100 Sensor? <sp>{C,X}<cr><lf> Returns the active sensor as above.
101 Sensor.acc? <sp>{value}<cr><lf> Returns the accuracy of the sensor.
102
Sensorid? <sp>{Address}<sp>MENSOR
,<sp>600SNSR,<sp>{Serial
No.},V{V.VV}<cr><lf>
Returns the active sensor’s serial number and
firmware version.
103
Setpt {A value inside the upper/low-
er limits, and inside the range
of the active transducer and
turndown}
Sets the control setpoint for the active channel.
104
Setpt? <sp>{value}<cr><lf> Returns the control setpoint of the active chan-
nel in current units.
105
Span {desired pressure or ?} Sets span on active transducer (must be >50%
FS, and within 1% limit), or ? to clear previous
value. (See “Caldisable” command above ref#
19)
106 Span? <sp>{value}<cr><lf> Returns span scale factor for active transducer
and turndown.
107
Sparity {Even, ODD , NONE} Sets the serial parity.
108 Sparity? {<sp>{string}<cr><lf> Returns the serial parity.
109 Srqmask Stable, Error or both Sets the CPC 6000 to issue a service request
(SRQ) over the IEEE-488 when the pressure
control is stable, or an error occurs. These are
80 hex and 40 hex respectively.
110 Srqmask? <sp>{string}<cr><lf> Returns “stable”, “error” or “error, stable” de-
pending on the SRQ.
111 Sstop {1 or 2} Sets the serial stop bits.
112
Sstop? <sp>{value}<cr><lf> Returns the serial stop bits.
113
Stable? <sp>{Yes or No}<cr><lf> Returns YES if instrument is stable or NO.
114
Stabledelay {0 to 65535} Sets the stable time to the number of seconds
specified.
115
Stabledelay? <sp>{value}<cr><lf> Returns the stable time.
116 Stabletime {0 to 65535} Sets the stable time to the number of seconds
specified.
117
Stabletime? <sp>{value}<cr><lf> Returns the stable time.