Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000 91
To get to the Calibration screen press [Setup], [Sensor], and then press [Calibrate].
10.6 Password
Autozero does not require a password, however one is needed to change any date of calibration, Zero,
Span, or to change the system password. The password installed at the factory was 1 2 3 4 5 6, but
the user can change this as described below.
10.6.1 Change Password
To change the current password:
1. Press the [Calibrate] key to see the Calibration Data screen similar to Figure 10.6.1A. Notice the
padlock.
Figure 10.6.1A - Calibration Data Screen (locked)
2. Touch the [Unlock] key to display the Enter Password window as shown in Figure 10.6.1B.
Figure 10.6.1B - ‘Enter Password’ Window