Wika T53 Manuel d'utilisateur Page 4

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 26
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 3
4
1.3 Physical Block (PA Slot 0) Parameter List, Profibus
PARAMETER
Rel.
Index Description Type Store Size R/W Min Max Default
ST_REV
1
Is incremented each time that there is a change in a
static parameter in the physical block.
Un-
signed
16
N 2 RO 0
TAG_DESC
2
Tag name of the block. This parameter must be unique
in the configuration.
OCTET_
STRING
SRC 32 R/W ‘ ‘
STRATEGY
3
This can be used to group a Function Block. It is a user
supplied parameter for identification purpose.
Un-
signed
16
SRC 2 R/W 0
ALERT_KEY
4 Current state of alarm blocks
Un-
signed
8
SRC 1 R/W 0
TARGET_MODE
5 Current desired mode of the block.
Un-
signed
8
SRC 1 R/W -
MODE_BLK
6
A block has static block parameters, that are not
changed by the process. Values are assigned to this
parameter during the configuration or optimisation. The
value of ST_REV must increase by 1 after every change
of a static block parameter. This provides a check of
the parameter revision.
DS-37 D 3 RO
Block
specific
ALARM_SUM
7 Current state of the blocks alarms. DS-42 D 8 RO 0,0,0,0
SOFTWARE_REVISION
8 Software revision of the device
VISBLE_
STRING
Cst 16 RO
HARDWARE_REVISION
9 Physical revision of the device.
VISBLE_
STRING
Cst 16 RO
DEVICE_MAN_ID
10
WIKA manufacturer identification number.
Un-
signed
16
Cst 2 RO
0x006B
DEVICE_ID
11 Manufacturer device number
VISBLE_
STRING
Cst 16 RO
WIKA
5310
DEVICE_SER_NUM
12 Device serial number
VISBLE_
STRING
Cst 16 RO
DIAGNOSIS
13
Bit string indicating the diagnosis of the device. See
Diagnosis
OCTET_
STRING
D 4 RO
DIAGNOSIS_EXTENSION
14 Not used.
OCTET_
STRING
D 6 RO
DIAGNOSIS_MASK
15 Not used.
OCTET_
STRING
Cst 4 RO
DIAGNOSIS_MASK_EXTENSION
16 Not used.
OCTET_
STRING
Cst 6 RO
DEVICE_CERTIFICATION
17 PA device certification
VISBLE_
STRING
Cst 32 RO
WRITE_LOCKING
18
If Locked, no writes from anywhere are allowed, except
to clear WRITE_LOCK. Cyclic block inputs will continue
to be updated.
Un-
signed
16
N 2 R/W
FACTORY_RESET
19
Factory reset:
1: Restart with default
2506: Restart processor
2712: Recover default address to the device
Un-
signed
16
SRC 2 R/W
DESCRIPTOR
20
It is a user supplied description of the block in the
application.
OCTET_
STRING
SRC 32 R/W
DEVICE_MESSAGE
21
It is a user supplied Message of the block in the appli
-
cation..
OCTET_
STRING
SRC 32 R/W
DEVICE_INSTAL_DATE
22 Date of the device installation.
OCTET_
STRING
SRC 16 R/W
LOCAL_OP_ENA
23 Not Used.
Un-
signed
8
N 1 R/W 1
IDENT_NUMBER_SELECT
24
0: Profile specific Ident_Num
1: Manufacture specific Ident_Number
2: Manufacture specific Ident_Number of V2.0
3: Ident_Number of Multi_Variable device
Un-
signed
8
SRC 1 R/W
HW_WRITE_PROTECTION
25 Unimplemented
RESERVED
26-32 Reserved to PNO (PROFIBUS Nutzerorganisation)
Vue de la page 3
1 2 3 4 5 6 7 8 9 ... 25 26

Commentaires sur ces manuels

Pas de commentaire