WIKA Betriebsanleitung Hand-Held Thermometer
2146872 09/2004 GB/D
12
GB
WIKA Operating Instructions Hand-Held Thermometer
Hand-Held Thermometer
CTH 6200
If keys ´Set´ and ´Store´ are pressed simultaneously for more the 2
seconds, the settings are reset of the settings ex-works.
If the logger memory contains data already, the menus/parameters
marked with (*) can not be invoked! If these should be altered the
logger memory has to be cleared before! (see chapter 3)
2. Configuration
To change device settings, press key: ´Set/Menu´ for 2 seconds.
This will call the configuration menu (main display: "SEt").
Pressing key: ´Set/Menu´ changes between the menus, pressing
key: ´Tara´ jumps to the referring parameters, which can be
selected with key: ´Tara´.
The parameters can be changed with key: ´Min´ or ´Max´. Pressing
key: ´Set/Menu´ again jumps back to the main configuration menu
and saves the settings.
Key: ´Store/Quit´ finishes the configuration and returns to stand-
ard measuring operation.
2.1 (Unit) Temperature unit
The measured temperature can be displayed in °C or °F.
2.2 (rES) Display resolution
Standard setting: ´Auto´ ,i.e. the device automatically switches over
to the optimum resolution between 0.1 ° and 0.01 °.
If temperatures to be measured are near the switching threshold, a
fixed resolution may be better, e.g. for easy recording.
In such a case please select the optimum resolution manually.
2.3 (Lin E.751) Sensor curve
This function used the standard conversion of resistance to
temperature following EN60751 (Lin E.751).