Wika IUT-11 Manuel d'utilisateur Page 64

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 78
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 63
Universal IS Pressure Transmitter UniTrans
Transmitter Operation with HART®-functionality
2266939.05 D/GB 06/2006
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: support-tronic@wika.de · www.wika.de
139
10.2.6 Submenu 'Parameter' - Evaluation
Evaluation:
Here the relation between height values and volume values is given in a graph.
Function:
Linear:
Here a linear relation between height and volume values is set.
Table:
The values of the table are set as linearisation graph between height and volume
value.
For liquid level measurements in tanks, values of height (level) must be entered and
a respective volume is assigned to each value. These value pairs are used to deter-
mine the linear relationship and to assign the 4 mA ... 20 mA output signal to values
of tank volume (P 0 and P 31 are fixed at 0 % and 100 %).