2266939.05 D/GB 06/2006
Universal IS Pressure Transmitter UniTrans
Transmitter Operation with HART®-functionality
Subject to change due to technical modifications. © Copyright WIKA Alexander Wiegand GmbH & Co. KG / Germany
WIKA Alexander Wiegand GmbH & Co. KG · Alexander-Wiegand-Str. · 63911 Klingenberg · (09372) 132 - 710 · Fax - 706 · E-mail: sup[email protected] · www.wika.de
140
Check lin. table:
The entered linearization table is subject to a plausibility check.
If wrong or incomplete values are entered the following error window pops up
.
Media density:
Here the density of the medium is given in g/cm³
The consequence of a correction or modification of the density value is that also the
respective values of the upper end of span are changed for the related measuring val-
ues (mm, m, inch, feet). In some cases after a media change (change of density) the
upper end of span must be recalibrated.