4. Design and function
4.1 Description
4. Design and function
Pressures p
1
and p
2
act on the media chambers ⊕ and ⊖, which are separated by
an elastic diaphragm (1).
The dierential pressure (Δp = p
1
- p
2
) leads to an axial deection of the
diaphragm against the measuring range springs (2).
The deection, which is proportional to the dierential pressure, is transmitted to
the movement (4) in the indicating case via a pressure-tight and low friction rocker
arm (3).
A magnet (5), xed to the rear of the movement, aects the electromagnetic eld
of the HALL sensor. The signal that results from this is converted to a standardised
current output signal by the signal conditioning board (6).
Overpressure safety is provided by metal bolsters (7) resting against the elastic
diaphragm.
14077606.01
⊖ ⊕
4.2 Scope of delivery