The channel is the input selector switch for function block (FB) of the
sensor. Within the function block, additional scalings (Out-Scale) are
carried out. In this menu item, the value r the function block is
selected:
l SV1 (Secondary Value 1):
- Pressure or height
l SV2 (Secondary Value 2):
- Percent
l PV (Primary Value):
- Linearised percentage value
Channel
PV lin. value
A pressure transmitter delivers the following me asured values:
l SV1 (Secondary Value 1): Pressure or height value before
adjustment
l SV2 (Secondary Value 2): Percentage value after the adjustment
l PV (Primary Value): Linearised percentage value
l PA-Out (value after passing the function block): PA output
l Temperature
In the menu item "Display" you can dene which measured value
should be presented on the display.
Displayed value
PA-Out
Probus transmits two values cyclically. The rst value is determined in
the menu item "Channel". The additional cyclical value is selected in
the menu item "Additional PA value".
With pressure transmitters the following values are available:
l SV1 (Secondary Value 1): Pressure or height value before
adjustment
l SV2 (Secondary Value 2): Percentage value after the adjustment
l PV (Primary Value): Linearised percentage value
Additional PA value
Channel
Display
Additional PA value
Indicating and adjustment module for IPT-1* sensors 21
5 Set up
31549-EN-081211
Vue de la page 20