6 Commissioning, operation
WIKA Operating Instruction, Model Pascal 100 and Pascal 100/IS 49
GB
The next step shows the span parameter (measuring range). In the default settings, the maximum
range is equal to the measuring range (span). If the span has to be redefined or reduced, it can be
changed by using a numeric keyboard. To access to the keyboard use the arrows displayed on the
right side of the screen. Change of the span is used during a calibration for comparison where a
relationship between a REF and a DUT channel has to be assigned.
Figure 53 mA range setup
Now it is possible to finalize the procedure pressing YES or come back to the previous step
pressing NO.
Figure 54 - Channel setup confirmation
Pressing YES, the channel is assigned to the selected parameter and the screen with four
channels is displayed again.
Figure 55 Main display with mA measurement