29WIKA operating instructions, model CPH65I0
14061681.02 03/2013 GB/D
GB
6. Commissioning, operation
The default resistance thermometer type is "PT100-385" (Pt100-385).
6.4.8 Damping
Damping can be turned ON or OFF using the "DAMPING" menu selection. As soon as
the damping function is switched ON, the calibrator displays the running average of the
last ten measurements. The calibrator makes approximately three readings per second.
6.5 Factory settings
The calibrator is delivered with ve commonly-used factory settings. These settings are
shown below.
Any of these settings can be changed and saved by the user.
Setting 1:
The upper display is set to [P1] mode and the middle is set to "mA"; the lower is o.