Automated Pressure Calibrator
CPC 6000
64 Mensor/WIKA Operating Instructions - CPC 6000
:MODE OFF/GAS/LIQUIT Sets the head correction mode.
:MEDIUM <n> Sets the density of the medium.
SYSTem
:DATE <i,i,i> Sets the system date (YY,MM,DD).
:TIME <i,i,i> Sets the system time (hh,mm,ss).
:ERRor[:NEXT]? Return: error#, “description”.
:KLOCk ON/OFF/1/0 Lock or unlock keyboard.
:PRESet Load known values.
:SAVe Saves settings to non-volatile memory.
:VERSion? Return of the SCPI-standard.
TEST
:ELECtronic? Returns electronics status.
:RELay<n>? Returns whether relay 1 to 3 is closed or not.
:RELay<n> ON/OFF Turns relay 1 to 3 on or off.
UNIT
:NAME<n>? Returns the name of the unit with index n.
:FACTOR<n>? Returns the conversion factor of the unit with index n.
index unit
0 bar
1 mbar
2 Pa
3 psi
4 atm
5 kp/cm2
6 lbf/ft2
7 kPa
8 cmH2O(4°C)
9 inH2O(4°C)
10 inH2O(60°F)
11 ftH2O(4°C)
12 μmHg(0°C)
13 mmHg(0°C)
14 cmHg(4°C)
15 inHg(0°C)
16 inHg(60°F)
17 - -
18 user
19 user
20 user
OUTPut
:STATe ON/OFF/1/0 Turns control mode on or off.
:STATe? Returns control mode “0” for off, ”1” for on.
:MODE MEASure/CONTrol/VENT Sets the operation mode.
:MODE? Returns current operation mode MEASURE”,
CONTROL”or “VENT”.
:STABLE? Returns 1 if in control mode and stable.