Automated Pressure Calibrator
CPC 6000
Mensor/WIKA Operating Instructions - CPC 6000 65
:AUTOvent ON/OFF/1/0 Sets autovent function on or off (automatically vent when
controlling at zero).
:AUTOvent? Returns status of Autovent function.
[SOURCE]
:PRESsure
[:LEVel]
[:IMMediate]
[:AMPLitude] <n> Sets the control pressure.
[:AMPLitude]? Returns current control pressure value.
:SLEW <n> Sets the slew rate in %FS/s.
:SLEW? Returns the current slew value in %FS/s.
:TOLerance <n> Sets the tolerance for stable indication in %FS.
:TOLerance? Returns the tolerance for stable indication in %FS.
CALCulate
:LIMit
:LOWer <n> Sets minimum control limit.
:LOWer? Returns minimum control limit.
:UPPer <n> Sets maximum control limit.
:UPPer? Returns maximum control limit.
7.4.7.3.1 Output Formats
1. <sp>pressure value<cr><lf>
2. <sp>pressure, units number, mode<cr><lf>
3. <sp>pressure, pressure rate<cr><lf>
4. <sp>pressure, minimum peak, maximum peak<cr><lf>
5. <sp>pressure, active sensor (P or S) active turndown (1-4)<cr><lf>
6. <sp>pressure, control point, “stable” or “slewing”<cr><lf>
7. <sp>pressure, “no barometer” or baro reading<cr><lf>
7.4.7.3.2 Error Messages and Error Codes
A recognized command is displayed at the device in the headline with “Remote...”. In case of error
“Remote error...” is displayed. A maximum of 100 errors are stored and can be retrieved successfully.
Table 7.4.7.3.2 - SCPI Command Error Messages and Error Codes
Code Error String Returned
0 No error
-101 Undefined character
-102 Syntax error
-103 Undefined separator
-104 Parameter data type error
-109 Parameter missing
-110 Undefined header
-113 Undefined command
-114 Parameter out of range