GB
3211517.0707/2010GB/D
24 WIKAoperatinginstructionstemperaturetransmitterT12
WARNING!
Anyexistingconnectionbetweenthepositiveandminus
terminalstoanyevaluationsystemmustbedisconnected
inordertoperformtheconguration.
Using the cable "RJ45 / RJ45"
Asanalternative,quickandeasyconnectionispossibleusing
the"RJ45/RJ45"cableincludedinthesuppliedconguration
set(onlyforT12.30).TodothistheprotectiveRJ45plug,whichis
deliveredinsertedinthefrontoftheunit,mustberemoved.Inits
place,theRJ45/RJ45cableleadingtotheprogrammingadapter
(programmingunit)mustbeinserted.Inthisway,thetransmitter
isconnectedtotheadapterwhiletheconnectiontotheevaluation
systemremainsinplaceviatheplus/minuswire.Inthiscase,it
isnotnecessarytodisconnecttheexistingwiringofthetransmit-
ter(positiveandnegativeterminals),sincethedisconnectionis
automaticallytriggeredbyswitchingtheRJ45plugs.
WARNING!
Afterconguration,theprotectiveRJ45plugmustbe
replacedintotheconnecteronthetransmitter(otherwise
thepositiveandnegativeterminalswillnotwork).
6. Commissioning, operation