30 WIKAoperatinginstructionstemperaturetransmitterT12
nAapplications:
Anexternaltransientprotectionofmax.140%ofsupplyvoltage
connectedtoterminals⊕and⊖shallbeprovidedbytheenduser
intheenduseapplication.
Operation in Zone 0:
Thetemperaturetransmittermayonlybeoperatedinareasthat
requireapparatusofcategory1whenfollowingatmospheric
conditionsexist:
Temperature:-20...+60°C
Pressure: 0.8...1.1bar
Operation in Zone 1 and Zone 2:
Accordingtothetemperatureclass,thesetransmittersmaybeused
onlyinthefollowingambienttemperatureranges:
Permissibleambienttemperatures
Headmountingversion
Model T12.10.xx9
II3GExnL
II3GExnA
II3GExic
T4:-40°C≤Ta≤+85°C
T5:-40°C≤Ta≤+75°C
T6:-40°C≤Ta≤+60°C
Railmountingversion
Model T12.30.xx9
II3GExnL
II3GExnA
II3GExic
T4:-20°C≤Ta≤+70°C
T5:-20°C≤Ta≤+70°C
T6:-20°C≤Ta≤+60°C
8.3 Safety-technical maximum values
8.3.1 Power supply / 4 ... 20 mA loop
Thefollowingsafety-relevantmaximumvaluesmustnotbeexceeded:
ModelT12.x0.xx9
II3GExnL
II3GExnA
II3GExic
Voltage: U
i
=DC36V