Wika CPC2000 Manuel d'utilisateur Page 20

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 132
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 19
20 WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
6.3 Operating mode 'AUTO'
In
AUTO
mode, a pressure prole can be stored. This is especially useful when multiple
pressure sensors or pressure switches are to be calibrated with the same values. The
mode also enables a dened prole to be traversed several times in succession.
Setting
Press the MENU button. Conrm the menu item MODE with the SELECT button and
press the
/ button until AUTO appears. Conrm with SELECT.
For the AUTO mode, the submenu points RANGE, UNIT and STEPS are not relevant.
Go to the SETTINGS menu item with the / buttons and conrm with the
SELECT button. The submenu item zero point adjustment is not relevant.
In the submenu, go to the AUTO mode item with the UP/DOWN buttons and conrm
with SELECT.
RANGE: Pressure range that will be covered
UNIT: Pressure unit
STEPS
: The number of steps which will be taken to increase and decrease the
pressure. These increments are calculated by the controller itself.
In the submenu point, CONFIGURATION 1, further settings can be dened for the
AUTO
mode.