21WIKA operating instructions, model CPC2000
11155094.02 07/2012 GB/D/F/E
GB
6. Commissioning, operation
The gure shows the ow chart for the AUTO mode. Based on the table, one will know
which cycles are being carried out and the meaning of each.
Setting
CYCLES: Here the number of cycles to be carried out in AUTO mode will be set.
T START: Delay time to vent the sensor
T STOP: Delay time before reaching the maximum dened pressure
T HOLD: Hold time until a new pressure value is reached
T PAUSE: Delay time between two cycles
AUTO ZERO: Automatic zero point adjustment at the completion of each cycle
To exit the submenu, press the MENU button.
Operation
On leaving the menu, the calibrator will revert to 'Standby status'. All information about the
AUTO mode will be shown on the display.
Operating mode
First of ten cycles Phase 1: It has been
vented (Pressure = 0)
percentage display of the current set
pressure
Pressure unit
Set pressure
Start
1 2 4 5 7 9 10 9 10 9 10 9 10 13 15 16 15 16 15 16 15 16 18
Next cycle
End of cycle