30 WIKA operating instructions, models CTH6300 and CTH6500
2079988.02 04/2013 GB/D
GB
6. Commissioning, operation
6.4.5 Activating/deactivating measuring channels (only for 2-channel
instruments) [Chnl]
Chnl = Channel = select
Press the ENTER/MENU key and select Chnl using the arrow keys ▲▼. Then press
ENTER/MENU again. A small 1 appears on the left side of the display, it indicates the
channel. Use the arrow keys ▲▼ to select the channel to be activated or deactivated.
Conrm selection by pressing ENTER/MENU. Use the arrow keys ▲▼ now to activate
on or deactivated o the set channel.
Use ESC to return to the measuring mode.
Alternative: keep the HOLD/MAX/MIN/AVE key pressed for 2 seconds; this deactivates
or activates channel 2.
At least one channel is always active!
6.4.6 Area entry for volume ow [ArEA]
The area dimensions can be entered only on instruments for ow measurement.
Press the ENTER/MENU key and select ArEA using the arrow keys ▲▼, then press the
ENTER/MENU key again. A small 1 appears on the left side of the display, it indicates
the channel. Use the arrow keys ▲▼ to select the channel. Conrm selection by
pressing ENTER/MENU. Now use the arrow keys ▲▼ to select the measuring unit:
c = centimeter²
m = meter²
Conrm selection by pressing ENTER/MENU. 00.00 is displayed now in the lower line
of the display. Change the blinking number by means of the arrow key ▲. Press the
arrow key ▼ to go to the next position. Conrm the input by pressing ENTER/MENU.
Press ESC to return to the measuring mode.
MENU
Unit Prob Lin2 CAL Chnl ArEA Lo6
MENU
Unit Prob Lin2 CAL Chnl ArEA Lo6